La relevancia del error técnico de medición en antropometría: Un estudio piloto del proyecto NutriFunction

Autores/as

  • Micaela Cunha Rodrigues FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal. LAETA-INEGI / FEUP, Associated Laboratory of Energy, Transports and Aerospace, Institute of Science and Innovation in Mechanical and Industrial Engineering, Faculty of Engineering of the University of Porto, Porto, Portugal. https://orcid.org/0000-0002-7140-3339
  • Rita Guerra LAETA-INEGI / FEUP, Associated Laboratory of Energy, Transports and Aerospace, Institute of Science and Innovation in Mechanical and Industrial Engineering, Faculty of Engineering of the University of Porto, Porto, Portugal. FP-I3ID, FP-BHS, Faculty of Health Sciences, University Fernando Pessoa, Porto, Portugal. https://orcid.org/0000-0003-4213-8529
  • Adriana Souza FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal. https://orcid.org/0000-0002-7938-7695
  • Rui Valdiviesso FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal. RISE-health, MEDCIDS, Faculty of Medicine of the University of Porto, Porto, Portugal. https://orcid.org/0000-0002-0124-5427
  • Bruna Rosa FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal. https://orcid.org/0009-0001-3074-329X
  • Mónica Rodrigues FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal. https://orcid.org/0000-0002-7932-9460
  • Maria Luisa Lucena FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal.
  • Ana S. Sousa FP-I3ID, FP-BHS, Faculty of Health Sciences, University Fernando Pessoa, Porto, Portugal.
  • Joana Mendes FP-I3ID, FP-BHS, Faculty of Health Sciences, University Fernando Pessoa, Porto, Portugal.
  • Ana Rita Sousa Santos University Institute of Health Sciences (IUCS - CESPU), Gandra, Portugal.
  • Nuno Borges FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal. RISE-health, MEDCIDS, Faculty of Medicine of the University of Porto, Porto, Portugal.
  • Teresa Amaral FCNAUP, Faculty of Nutrition and Food Sciences of the University of Porto, Porto, Portugal. LAETA-INEGI / FEUP, Associated Laboratory of Energy, Transports and Aerospace, Institute of Science and Innovation in Mechanical and Industrial Engineering, Faculty of Engineering of the University of Porto, Porto, Portugal. https://orcid.org/0000-0002-3998-6730

DOI:

https://doi.org/10.14306/renhyd.28.1.2312

Palabras clave:

Antropometría, Confiabilidad, Error de medición intra-observador, Error de medición inter-observador, Error técnico de medición

Resumen

Introducción. La antropometría es crucial en la evaluación nutricional, especialmente en entornos clínicos e investigativos donde la falta de precisión afecta la confiabilidad de los datos y la atención al paciente. Sin embargo, la precisión de estas mediciones suele pasarse por alto al evaluar el estado nutricional y los riesgos de salud. Este estudio tuvo como objetivo evaluar el error técnico de medición (TEM) y el error intra e inter-observador en las mediciones antropométricas realizadas por el equipo de NutriFunction, un estudio multicéntrico que incluye pacientes hospitalizados y una muestra de individuos que viven en la comunidad.

Metodología. Seis entrevistadores capacitados evaluaron doce indicadores antropométricos: peso corporal, altura corporal y de la rodilla, longitudes de la mano y del dedo medio, anchura de la mano, circunferencias del brazo medio, cintura y pantorrilla, grosor del tríceps, pantorrilla y aductor del pulgar, siguiendo los protocolos de ISAK y otras técnicas estandarizadas. Se calcularon el TEM, el %TEM y el coeficiente de fiabilidad (R).

Resultados. La mayoría de los resultados fueron aceptables para antropometristas experimentados, excepto la anchura de la mano (%TEM=1.2) y la altura de la rodilla (%TEM=1.2) para el antropometrista #4, y la longitud del dedo medio (%TEM=1.3) para el #5 en variabilidad intra-observador (R: 0.6-1.0). La variabilidad inter-observador presentó %TEM inaceptables para la longitud de la mano y el grosor del aductor del pulgar en varios antropometristas (R: 0.3-1.0). La variabilidad intra-observador fue aceptable para principiantes, pero la inter-observador no lo fue para la longitud de la mano y el grosor del aductor del pulgar en el antropometrista #2.

Conclusiones. La mayoría de las mediciones mostraron TEM aceptables, indicando precisión suficiente para evaluación y diagnóstico clínico. El antropometrista #2 necesita entrenamiento adicional antes de medir la longitud de la mano y el grosor del aductor del pulgar. El entrenamiento regular es esencial para minimizar errores y mantener la calidad y utilidad de los datos para evaluación, diagnóstico y tratamiento.  

Financiamiento: Beca FCT (2023.01790.BD) y European Regional Development Fund (NORTE-01-0145-FEDER-000039).

Citas

Gibson RS. Principles of Nutritional Assessment: Oxford University Press; 2005.

Jelliffe DB, Jelliffe EFP. Community Nutritional Assessment: With Special Reference to Less Technically Developed Countries: Oxford University Press; 1989.

Ulijaszek SJ, Kerr DA. Anthropometric measurement error and the assessment of nutritional status. Br J Nutr. 1999;82(3):165-77.

Holmes CJ, Racette SB. The Utility of Body Composition Assessment in Nutrition and Clinical Practice: An Overview of Current Methodology. Nutrients. 2021;13(8):2493. 10.3390/nu13082493

Müller MJ, Braun W, Pourhassan M, Geisler C, Bosy-Westphal A. Application of standards and models in body composition analysis. Proc Nutr Soc. 2016;75(2):181-7. 10.1017/S0029665115004206

Norton K, Olds T. Anthropometrica. Sydney: University of New South Wales Press1996.

Hume P, Marfell-Jones M. The importance of accurate site location for skinfold measurement. J Sports Sci. 2008;26(12):1333-40. 10.1080/02640410802165707

Nádas J, Putz Z, Kolev G, Nagy S, Jermendy G. Intraobserver and interobserver variability of measuring waist circumference. Med Sci Monit. 2008;14(1):Cr15-8.

Scafoglieri A, Clarys JP, Cattrysse E, Bautmans I. Use of anthropometry for the prediction of regional body tissue distribution in adults: benefits and limitations in clinical practice. Aging Dis. 2014;5(6):373-93. 10.14366/AD.2014.0500373

Zubir S, Ikhwan MN, Abd. Rahman S, Hamadan H, Salim N, Haji Baruji THME, et al. Technical Error Measurements, Reliability, and Validity of Customized Anthropometric Grid. Journal of Occupational Safety and Health. 2023;20:51-8.

Hardy J, Kuter H, Campbell M, Canoy D. Reliability of anthropometric measurements in children with special needs. Arch Dis Child. 2018;103(8):757-62. 10.1136/archdischild-2017-314243

Carsley S, Parkin PC, Tu K, Pullenayegum E, Persaud N, Maguire JL, et al. Reliability of routinely collected anthropometric measurements in primary care. BMC Medical Research Methodology. 2019;19(1):84. 10.1186/s12874-019-0726-8

Madden AM, Smith S. Body composition and morphological assessment of nutritional status in adults: a review of anthropometric variables. J Hum Nutr Diet. 2016;29(1):7-25. 10.1111/jhn.12278

Frisancho AR. Anthropometric standards for the assessment of growth and nutritional status. Ann Arbor: University of Michigan Press; 1990. Available from: http://catalog.hathitrust.org/api/volumes/oclc/20932481.html.

Lohman TG, Roche AF, Martorell R. Anthropometric standardization reference manual. Champaign, IL: Human Kinetics Books; 1988.

Ulijaszek SJ, Mascie-Taylor CGN. Anthropometry: The Individual and the Population: Cambridge University Press; 1994.

Marks GC, Habicht JP, Mueller WH. Reliability, dependability, and precision of anthropometric measurements. The Second National Health and Nutrition Examination Survey 1976-1980. Am J Epidemiol. 1989;130(3):578-87.

Arroyo M, Freire M, Alday L, Pablo AM. Intraobserver error associated with anthropometric measurements made by dietitians. Nutr Hosp. 2010;25:1053-6.

Klipstein-Grobusch K, Georg T, Boeing H. Interviewer variability in anthropometric measurements and estimates of body composition. International journal of epidemiology. 1997;26 Suppl 1:S174-80.

F. Esparza-Ros, R. Vaqueri-Cristóbal, Marfell-Jones. M. International standards for anthropometric assessment, 2019th ed. Murcia, Spain: International Society for the Advancement of Kinanthropometry. 2019.

Lameu EB, Gerude MF, Campos AC, Luiz RR. The thickness of the adductor pollicis muscle reflects the muscle compartment and may be used as a new anthropometric parameter for nutritional assessment. Curr Opin Clin Nutr Metab Care. 2004;7(3):293-301.

ISO7250-1:2017. Basic Human Body Measurements for Technological Design—Part 1: Body Measurement Definitions and Landmarks. ISO/TC 159/SC 3 Anthropometry and Biomechanics.

Guerra RS, Fonseca I, Pichel F, Restivo MT, Amaral TF. Hand length as an alternative measurement of height. Eur J Clin Nutr. 2014;68(2):229-33.

Lee RD, Nieman DC. Nutritional assessment: McGraw-Hill Boston, MA, USA:; 2007.

Pereira PML, Neves FS, Bastos MG, Cândido APC. Adductor Pollicis Muscle Thickness for nutritional assessment: a systematic review. Rev Bras Enferm. 2018;71(6):3093-102.

Himes JH. Reliability of anthropometric methods and replicate measurements. Am J Phys Anthropol. 1989;79(1):77-80.

Gore C, Norton K, Olds T, Whittingham N, Birchall K, Clough M ea. Accreditation in anthropometry: an Australian model. In: Norton K, Olds T, editors, Anthropometrica. Sydney: University of New South Wales Press. 1996:p. 395–411.

Oliveira T, Oliveira G, Ornellas J, Oliveira F. Technical error of measurement in anthropometry (English version). Revista Brasileira de Medicina do Esporte. 2005;11:81-5.

Gonzalez MC, Duarte RR, Budziareck MB. Adductor pollicis muscle: reference values of its thickness in a healthy population. Clinical nutrition. 2010;29 2:268-71.

Weinberg SM, Scott NM, Neiswanger K, Marazita ML. Intraobserver error associated with measurements of the hand. Am J Hum Biol. 2005;17(3):368-71.

Kouchi M, Mochimaru M, Tsuzuki K, Yokoi T. Interobserver errors in anthropometry. J Hum Ergol (Tokyo). 1999;28(1-2):15-24.

Publicado

2025-03-15

Cómo citar

Cunha Rodrigues, M., Guerra, R., Souza, A., Valdiviesso, R., Rosa, B., Rodrigues, M., Lucena, M. L., Sousa, A. S., Mendes, J., Sousa Santos, A. R., Borges, N., & Amaral, T. (2025). La relevancia del error técnico de medición en antropometría: Un estudio piloto del proyecto NutriFunction . Revista Española De Nutrición Humana Y Dietética. https://doi.org/10.14306/renhyd.28.1.2312

Número

Sección

Artículos de investigación